Technical characteristics:
- Operating mode: semi-automatic
- Signal format: bi-phase square wave (truncated exponential); compensated impedance.
- Energy:
- 200J fixed
- Adult: rated 150J, 50ohm load.
- Infant/child: nominal 50J, 50ohm load.
- Energy delivered accuracy: within ± 3% (200J, 50ohm load).
- Less than 20 seconds pause between two shocks.
- Voice guidance: guides you through the use with voice assistance
- Charging control: automatic with software control (Arrhythmia Detection System and Charging control).
- Charging time from "Shock recommended": usually less than 10 seconds
- Controls: Shock button, information button, on/off button.
- Feedback: 7 LEDs (different colours), information button.
- Ready to charge feedback:
- For example, with text feedback ("Press the flashing orange button now")
- Flashing backlight on the Shock button
- Deactivation: when charging, the device will automatically deactivate itself if:
- The patient's heart rhythm goes into a non-shockable state, or
- The shock button is not used within 15 seconds after charging, or
- The on/off button is pressed to turn the device off, or
- The defibrillator electrodes are removed from the patient or the connector is disconnected.
- Shock: by pressing the shock button when the device is charged.
- Battery: 12V, 4.2 Ah, Lithium manganese dioxide, non-rechargeable long-life primary cell
- Battery capacity: minimum 120 discharges or 4 hours of operation.
Battery capacity 120 volts per battery (120 volts per charge)
Daily: checks internal circuits, signal generator, battery capacity, and software
Battery: Comprehensive self-test when battery is inserted and user-interactive test to check the device is ready for use
- Data recording and transmission:
Infra port: Wireless data transmission can be established between the PC and the device using the infrared port (Software and Irda cable optional).
Data storage: first 40 minutes of ECG scan and the complete accident history and decisions made during the analysis.
- Technical features:
- Operating Environment:
- Temperature: 0°C to 40°C (32°F to 110°F)
- Humidity: 5 % to 95 % (non-condensing)
- Storage environment:
- Temperature: 0 °C to 40 °C (32 °F to 110 °F)
- Humidity: 5 % to 95 % (non-condensing)
- Size: 220x281x82 mm
- Weight: 2.2 Kg (4.48 lbs) with battery.
- Mechanical tests (impact and breakage test): meets IEC 60601-1 clause 21 (Mechanical strength)
- Vibration: meets EN 1789 accidental, road rescue requirements in working and standby conditions.
- Water resistance: complies with IEC 60529: IP54
- ESD (electrostatic discharge): complies with IEC 61000-4-2:2001
- EMI (radiated) compliance: IEC 60601-1-2 with limit value, EN 55011:1998+ A1:1999 +A2:2002, Group 1, Class B
- EMI (resistance) compliance: IEC 60601-1-2 limit, EN 61000-4-3:2001 Level 3 (10V/m 80MHz to 2500MHz)